How to Demonstrate the Error Location Capture Measurement
To demonstrate the Error Location Capture measurement:
1
Switch to the
Pattern
panel and press the
Open
button. Load a Fiber Channel
random data pattern from the demo patterns (Demo > Fiber Channel > RPAT.ptrn).
Press the
Properties
button and set the
Pattern Type
to
Alternate
. To display
both halves of the alternate pattern, press the
Alt pat view
button.
Ensure also that the error detector tracks the pattern generator.
2
Insert two errors into pattern B. To do so, edit bits 80 and 100 of pattern B in
overwrite mode.
3
Write the modified pattern to the pattern generator and the error detector by
pressing the
ToPG&ED
button.
4
For the pattern generator setup you need to specify the
Alternate Pattern and
Aux In
. Set the
Internal Alternate Pattern Mode
to
Single Shot B
and disable
Aux In
.
5
Set up the error detector so that the
Clock Setup
is set to
Clock Data
Recovery
to get the error detector's clock from the incoming data stream.
6
Switch to the
Analysis
panel and then press the
Error Location Capture
icon.
7
Press the
Start
button to execute the measurement.
Status information is displayed in the lower section of the display.
8
Press
Insert B
button.
The two captured errors are illuminated red in the captured pattern grid. In the
result list, the
1st Bit-Error Location Capture
indicates an error in bit 80. Use
the error navigation buttons (e.g.
Next Error
,
Prev. Error
) to display further bit
errors in the captured pattern.
Error Location Capture - Reference
The Error Location Capture measurement compares the expected pattern and the
pattern that was actually received, and shows the received pattern with the errored
bits marked.
Elements of the User Interface
The following buttons help you to analyze the results:
•
Hex / Bin
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
305
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...