4
Setting up the Pattern Generator
Setting up the Pattern Generator - Concepts
The Serial BERT's pattern generator generates an output signal based on a data
pattern. It has the following possibilities for generating an output signal:
• Providing a wide range of clock frequencies
You can use the pattern generator's internal clock or an external clock for
defining the frequency of the outgoing stream.
• Distorting the signal by adding jitter
You can connect an external delay control device (for example a function
generator) to add jitter to the generated signal.
• Adding errors to the output stream
The Serial BERT can be set up to insert errors into the outgoing stream either
internally, according to an external signal, or manually (from the operator).
• Run-time switching between two patterns
You can set up two patterns and switch between them during runtime either
automatically, according to an external signal, or manually.
• Suppressing the output stream
The output signal can be suppressed according to an external signal.
The pattern generator also provides output ports that let you connect an external
instrument, such as an oscilloscope.
Agilent J-BERT N4903B High-Performance Serial BERT
User Guide
s
Agilent Technologies
121
S
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...