4
From the
Navigation Menu
, select
External Instrument(s)
and then click
Config
sub-menu. The
Config
window shows a list of all instruments connected
to the Serial BERT. Confirm the presence of N4916A enrty in that list.
5
Corrosponding to N4916A entry, select the
De-Emphasis
function check box
from the given options under
Enable Function
column. It opens the
De-
emphasis Signal Converter Connection
dialog.
6
Click
Enable
button. It enables N4916A De-emphasis signal converter that is
connected between data out of Serial BERT and the DUT.
Once the de-emphasis function is enabled you can see the de-emphasis
parameters in the PG Output screen.
7
Again, from the
Config
window, corrosponding to N4916A, clear the
De-
Emphasis
function check box. It will disable the de-emphasis function.
If you could proceed until here, the connection of the user software to the N4916A
is ok.
Check the Output of the N4916A Using a Scope
1
Set the Data offset to 0 V and the amplitude to 1 V. Set the De-Emphasis to 6
dB.
2
Connect the Output of the N4916A to an oscilloscope.
3
Connect the trigger output of the Serial BERT to the trigger input of the
oscilloscope.
9
Solving Problems
496
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...