Depending on the quality and characteristics of the eye opening, the
resulting 50% threshold may deviate from the decision threshold defined in
the loaded setting.
•
Number of Valid Points
Change the
Number of Valid Points
, if you wish to use less or more
measurement points for the measurement.
•
Relative Time and Voltage
In the table at the bottom of the
Parameters
tab you can change measurement
points. Enter the
Relative Time
and
Voltage
according to the above settings.
Pass/Fail Tab
The
Pass/Fail
tab of the
Properties
dialog box allows you to specify the criteria
to decide whether the DUT passes or fails the test. You can change pass/fail
criteria without rerunning a test. The software only uses the criteria to rate the
results of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails one or more of the criteria.
The Fast Eye Mask measurement has only one pass/fail condition: The bit error
rate. If desired, enable the
BER Threshold
and set a suitable threshold.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
313
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...