3
Set the
Number of Steps
between start and stop.
The
Number of Steps
yields a set of logarithmically equidistant frequencies.
In
Manual
mode, the automatically calculated jitter frequencies can be edited. You
can also enter your own comma-separated list of frequencies.
The
Sequence Advance
checkbox advances the PG sequence during jitter
tolerance characterization. It will be enabled only if the sequence is send to the
PG.
The
Show Compliance Curve
checkbox enables the compliance curve on the Jitter
Tolerance Characterization graph. This feature is available only if the compliance
jitter measurement option is installed. The standard is taken from the compliance
measurement.
Setting the BER Limit
1
Set the
Target Bit Error Ratio
.
This defines whether a measured point is flagged green or red.
2
Choose the
Verification Method
.
Setting a
Confidence Level
is generally recommended. See also
.
If you enable
Absolute Bits/Errors
, the measurement proceeds with the next
amplitude as soon as one of the two numbers is reached.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
439
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...