• Change the ‘centerspread’value: the bitrate remains unchanged, while the
upper and lower frequency changes according to the selected deviation. The
deviation value specifies ½ p-p value.
• Change the ‘downspread’ value: the upper frequency remains unchanged,
while the bitrate is adjusted. The deviation value specifies the p-p value.
For some setups, the I/Q modulator’s range limitations require to use centerspread
instead of downspread and adjust the bitrate (and deviation) accordingly. To ease
the handling of this setup, the SSC
Advanced Edit Panel
allows to edit the upper,
center, and lower frequency.
If the upper or lower frequency is change, the bitrate will be adjusted according to
the selected deviation. The center frequency corresponds to the configured bitrate.
The adjust buttons allow to adjust the corresponding frequency by the specified
ppm-value.
Please note that both the upper and lower frequency and the frequency adjustment
of all three frequencies are not reflected in the firmware. The GUI just calculates
the resulting bitrate and writes it to the firmware.
The
Deviation
of the clock rate. The range of SSC deviation are as following:
• Downspread with deviation of 0% … 0.5 %
• Centerspread with deviation of 0% … ± 0.5 %
• Upspread with deviation of 0% … 0.5 %
N O T E
The deviation in Upspread and Downspread is the p-p value while in Centerspread
it is 1/2p-p.
The deviation
Frequency
. The SCC provides the frequency range of 10 kHz – 100
kHz.
You can click on the corrosponding button to choose among Downspread,
Centerspread and Upspread deviation type. The green LED on the button indicates
the currently selected deviation type.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
417
Deviation
Frequency
Type
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...