Low Level
Specifies the lower limit of the measurement voltage range. Enter the value in
mV. This value should be slightly lower than the lowest expected signal voltage.
•
High Level
Specifies the upper limit of the measurement voltage range. Enter the value in
mV. This value should be slightly higher than the highest expected signal
voltage.
Pass/Fail Tab
The
Pass/Fail
tab of the
Properties
dialog allows you to specify the criteria to
decide whether the DUT passes or fails the test. You can change pass/fail criteria
without rerunning the test. The software only uses the criteria to rate the results
of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails one or more of the criteria.
You can set several pass/fail limits. For an explanation of the
Eye Opening Pass/
Fail
criteria, see
“Measurement Parameters” on page 301
6
Advanced Analysis
296
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...