7
Again, from the
Config
window, corresponding to M8061A, clear the
Mux with
Deemphasis
function check box. It will disable the multiplexer function, and
the PG Output screen will allow to control the output levels of Data and Aux
Data at the connectors of the PG again.
If you could proceed until here, you have successfully connected the application
software to the M8061A.
Settings for Multiplexer Tests
The following are the settings to test the output of the M8061A 28 Gb/s Multiplexer
with Deemphasis:
• Preset Instrument State.
• Set the PRBS 2^7-1.
• In the
PG-Bit Rate Setup
window, set the
Data Rate
to 3 Gb/s.
• Enable
Mux with DeEmphasis
function in the
Config
window.
• In the
M8061A
window, set the
Data Offset
to 0 V.
• In the
M8061A
window, set
Data Amplitude
to 400 mV.
Check the Output of the M8061A Using a Scope
The approach shown here is checking the output signal using an oscilloscope.
1
Connect the Outputs of the M8061A to an oscilloscope.
2
Connect the Trigger Output of the pattern generator to the trigger input of the
oscilloscope.
3
Terminate M8061A Data /Out with 50 Ohm.
4
Terminate PG Trigger-/Out with 50 Ohm.
5
Terminate PG Data /Out and Aux /Data Out with 50 Ohm.
Then the oscilloscope should show an output eye like the following:
If you see a display like above everything is OK.
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
521
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...