Green points (BER below limit) and red crosses (BER above limit) are used to
display the results.
After a measurement has exceeded the target BER or when moving to the next
frequency point, a measurement settling or relax time can be applied. This allows
the DUT to settle after changes and before the next measurement; this is important
because some devices may need time to recover from error conditions caused by
jitter exceeding the margin during the search for the threshold.
The maximum jitter amplitude the Serial BERT can generate is also frequency
dependent, as illustrated below.
Sinusoidal jitter (SJ) and periodic jitter (PJ) with sinusoidal characteristic are used
for the test. Whether SJ or PJ is used depends on the jitter frequency. SJ supports
wide amplitude variations up to some MHz. PJ supports narrow amplitude
variations over a wide frequency range. The measurement switches automatically
between the two sources.
8
Jitter Tolerance Tests
434
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...