User-Defined Standards
You can create your own standard in a text file.
A user-defined standard consists of a paired list of numbers.
Each pair specifies a jitter frequency (Hz) and the associated jitter amplitude (UI),
for example:
2e3, 110<cr><lf>
3.24e4, 50<cr><lf>
71.234e4, 10<cr><lf>
10.654e6, 0.2<cr><lf>
18.333e6, 0.2<cr><lf>
<cr><lf> stands for <carriage return><line feed> or just <Return>.
N O T E
The list must be sorted according to the frequencies in ascending order. The jitter
amplitude for a given frequency must not exceed the instrument’s capability.
The list of predefined standards has a
Browse
option that allows you to identify
and load such a text file.
If you have selected a text file, you can append it to the list of predefined standards.
You can assign a name to your standard and add a description. Then, your standard
appears directly in the list like the standards provided by Agilent, and you don’t
have to browse any more.
Your standard is stored in a file with the suffix .jcs. This is still a text file which can
be edited. The directory is C:\<instrument model>\JTolStandards.
User-defined standards are taken “as is”. That means, neither the bit rate nor any
other conditions are taken into account.
The instrument uses double logarithmic interpolation to generate the jitter
amplitudes between two successive amplitude values. This is indicated by linear
slopes in the double logarithmic graph.
Special Features
You can freely specify the jitter frequency range and the number of points to be
measured. You can also adjust the frequency range and the point positions to the
loaded standard.
In addition, you can add a margin to the standard.
8
Jitter Tolerance Tests
452
Agilent J-BERT N4903B High-Performance Serial BERT
User-Defined Standard Specifications
Using User-Defined Standards
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...