This displays the list of measured points. The list classifies the points with their
frequency, amplitude, and status. The status indicates whether the test passed or
failed at that point.
This contains the results as follows:
1
Test Configuration Details: gives the User Comments, Device Type, Last Test
Date, Model Number, Serial Number and Software Revision number.
2
Conditions: Pattern, Data Rate, Output Level, ISI Trace Number, Additional
Jitter, SSC, Target BER, Confidence Level, Relax Time and Margin.
3
Graph: Jitter Tolerance Graph.
4
Measured Points: List of points classified with their pass/fail status,
frequency, Periodic Jitter/Sinusoidal Jitter and Total Jitter without ISI.
5
Recommended Settings (only in the Compliance Report): Document, Note, Bit
Rate, Pattern, Level from Standard, Level at PG (recommended) and Additional
Jitter.
Explanation of the Results
Additionally to the graphical results, the measurement provides some information
below the graph. You can also export the results to a .txt file.
The following information helps you to interpret and document the test:
•
Result
of the test (passed or failed).
•
Date and Time
when the test was finished.
•
Margin
that was set for the test.
• Further test
Conditions
like the selected test pattern and additional jitter
components.
• Double-click
Modify Comment
, to enter a
Comment
for every test.
T I P
If not the whole information is displayed in the
Condition/Comment column, this
is indicated by "…". Double-click
Modify Comment to see the complete text in a
separate window.
If you want to use the measurement results with other applications, you can export
the data to a file via
Analysis > Export Data..
..
The contents of the resulting file may look as follows:
8
Jitter Tolerance Tests
466
Agilent J-BERT N4903B High-Performance Serial BERT
Select Point
HTML Report
Understanding the Results
Exporting Result Data
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...