You can enable the input after reducing the voltage to a safe level.
Delay Control Input - Procedures
N O T E
The following procedure does not apply to Serial BERT on which the calibrated and
integrated jitter injection option J10 is installed. For such instruments refer to
“Jitter Setup - Procedures” on page 407
.
To set up a jitter tolerance test on an instrument without option J10:
1
Press
PG Data Setup
in the
PG Setup
menu.
2
Disable the Delay Control In port by clearing the
Delay Control Input
checkbox.
3
Connect the device that supplies the jitter signal to the Delay Control In port.
4
Enable the port again by selecting the
Delay Control Input
checkbox.
T I P
While monitoring the BER you can use this checkbox to switch jitter on and off and
view the effect on the BER.
Delay Control Input - Reference
This checkbox in the
PG Data Setup
window is used to disable or enable the Delay
Control In port, where you can supply an external signal to add jitter to the output
signal.
4
Setting up the Pattern Generator
156
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...