• numbers of received bits and errors
• a confidence level
Additionally, a relax time can be specified.
Number of Received Bits or Errors
When numbers of received bits or errors are enabled, the measurement proceeds
with the next frequency as soon as one of the two numbers is reached.
But we are often dealing with very small tolerable bit error ratios, 10
–12
and below.
Measuring such a bit error ratio precisely (and just for one single point) takes time.
Therefore, the measurement provides the option to set a confidence level.
Confidence Level
The Jitter Tolerance Compliance measurement offers the option to set a level of
confidence. This approach is based on statistics.
Statistics tell us, for example, that if we receive 3 × 10
12
bits without any error, the
probability that the BER is below 10
–12
is higher than 95 %. If an error occurred,
more bits must be compared to achieve the same confidence level. For details see
“Explanation of the Fast Total Jitter Measurement” on page 246
.
The measurement proceeds with the next frequency and jitter amplitude as soon
as the desired level of confidence is reached. Setting a confidence level ensures
that neither too few nor too many bits are captured.
Relax Time
Before the measurement proceeds to the next frequency bin, a measurement
settling or relax time can be applied.
This allows the DUT to recover in case it has lost synchronization due to an
excessive high bit error rate. The next BER measurement starts after this time has
elapsed.
About Jitter Tolerance Standards
Every Jitter Tolerance Compliance test requires a standard. You can use one of the
predefined standards. Additionally, you can define and use your own "user-
defined" standard.
In terms of the Serial BERT, a standard is represented by a sequence of paired
numbers. Each pair consists of a frequency value (Hz) and the associated jitter
amplitude value (UI).
Also part of a standard specification are the standard’s name and descriptive text.
8
Jitter Tolerance Tests
450
Agilent J-BERT N4903B High-Performance Serial BERT
Содержание J-BERT N4903B
Страница 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Страница 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 486: ... Input Timing Setup 2 Input Timing Setup 3 9 Solving Problems 486 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 487: ... Input Timing Setup 4 Input Timing Setup 5 Solving Problems 9 Agilent J BERT N4903B High Performance Serial BERT 487 ...
Страница 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Страница 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...