Index
2.6 Triggering
Overview
The trigger function of the GS820 achieves various types of triggering through
combinations. The following three types of triggers are available.
Source Trigger
(See section 5.7 for the procedure)
This trigger is used to start a source action (see page 2-9). Select any of the following
trigger sources.
• External trigger
• Auxiliary trigger (rising edge) (Aux )
• Auxiliary trigger (falling edge) (Aux )
• Timer1
• Timer2
• Measurement end (MeasEnd)
Measurement Trigger
(See section 7.6 for the procedure)
This trigger is used to start a measurement action (see page 2-17). Select any of the
following trigger sources.
• Source change (SrcChg)
• Auxiliary trigger (rising edge) (Aux )
• Auxiliary trigger (falling edge) (Aux )
• Timer1
• Timer2
• Immediate (Imm)
• Sweep end (SwpEnd)
Sweep Start
(See section 5.6 for the procedure)
This trigger is used to start a sweep operation (see section 2.5, “Sweep”). Select any of
the following trigger sources.
• External sweep start
• Auxiliary trigger (rising edge) (Aux )
• Auxiliary trigger (falling edge) (Aux )
• Timer1
• Timer2
• Measurement end (MeasEnd)
2-25