16-15
IM 765601-01E
Communication Commands
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
16.2.4 Source Commands (SOURce Group)
[:CHANnel<n>]:SOURce:RESPonse/?
Function Sets the response mode (normal or stable) or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce:RESPonse
NORMal|STABle
NORMal
Selects normal.
STABle
Selects stable.
[:CHANnel<n>]:SOURce:RESPonse?
→ NORM Set to normal.
→ STAB
Set to stable.
Example
:SOUR:RESP NORM
:CHAN2:SOUR:RESP STAB
:CHAN1:SOUR:RESP?
[:CHANnel<n>]:SOURce:TRIGger/?
Function Sets the trigger source (external, auxiliary trigger,
Timer1, Timer2, or measurement end) or queries
the current setting.
Syntax
[:CHANnel<n>]:SOURce:TRIGger
EXTernal|AUXiliary|TIMer1|TIMer2|
SENSe
EXTernal
Selects external trigger.
AUXiliary
Selects auxiliary trigger.
TIMer1
Selects Timer1 (constant period).
TIMer2
Selects Timer2 (constant period).
SENSe
Selects measurement end.
[:CHANnel<n>]:SOURce:TRIGger?
→ EXT
Set to external start.
→ AUX
Set to auxiliary trigger.
→ TIM1
Set to Timer1.
→ TIM2
Set to Timer2.
→ SENS
Set to measurement end.
Example
:SOUR:TRIG TIM1
:CHAN2:SOUR:TRIG EXT
:CHAN1:SOUR:TRIG?
16.2 Commands
[:CHANnel<n>]:SOURce:TRIGge:
AUXiliary:POLarity/?
Function Sets the auxiliary trigger polarity (normal or
inverted) or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce:TRIGge:
AUXiliary:POLarity NORMal|INVerted
NORMal
Selects normal (falling edge).
INVerted
Selects inverted (rising edge).
[:CHANnel<n>]:SOURce:TRIGge:
AUXiliary:POLarity?
→ NORM Set to normal (falling edge).
→ INV
Set to inverted (rising edge).
Example
:SOUR:TRIG:AUX:POL NORM
:CHAN2:SOUR:TRIG:AUX:POL INV
:CHAN1:SOUR:TRIG:AUX:POL?
[:CHANnel<n>]:SOURce:FUNCtion/?
Function Sets the source function (voltage or current) or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce:FUNCtion
VOLTage|CURRent
VOLTage
Selects voltage.
CURRent
Selects current.
[:CHANnel<n>]:SOURce:FUNCtion?
→ VOLT
Set to voltage.
→ CURR
Set to current.
Example
:SOUR:FUNC VOLT
:CHAN2:SOUR:FUNC CURR
:CHAN1:SOUR:FUNC?
[:CHANnel<n>]:SOURce:SHAPe/?
Function Sets the source waveform (DC or pulse) or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce:SHAPe DC|PULSe
DC
Selects DC.
PULSe
Selects pulse.
[:CHANnel<n>]:SOURce:SHAPe?
→ DC
Set to DC.
→ PULS
Set to pulse.
Example
:SOUR:SHAP DC
:CHAN2:SOUR:SHAP PULS
:CHAN1:SOUR:SHAP?