![YOKOGAWA GS820 User Manual Download Page 58](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275058.webp)
Sweep Start Output, Trigger Output, and Auxiliary Trigger Output
(See sections 10.1 and 10.3 for the procedure)
The GS820 can output the sweep start, source trigger, and auxiliary trigger signals from
the SYNC OUT terminal (output terminal for synchronous operation) or the BNC output
terminal to the subsequent GS820s that are connected.
StartOut
A sweep busy for CH1 (set to low when sweep start is applied and set to high when the
sweep operation is completed).
TrigOut
A source busy for CH1 (set to low when a source trigger is applied and set to high when
the source action is completed).
AuxOut
An auxiliary trigger output for CH1. This signal is a logic OR of its own auxiliary trigger
input and a specified auxiliary trigger source signal.
Trigger Hold
The trigger hold function temporarily stops source triggers and measurement triggers.
A mask is applied when the HOLD key on the front panel is pressed or when the
“
:TRIGger:HOLD ON
” communication command is received, and the source action
and measurement action stop. To resume, press the HOLD key again or send the
“
:TRIGger:HOLD OFF
” communication command.
Sampling Error
The sampling error indicator on the front panel indicates that a trigger has been
discarded. The indicator illuminates if any of the errors below occurs on CH1 or CH2.
• A new source trigger occurred while a source action was in progress (Source Busy is
low).
• A new measurement trigger occurred while a measurement action was in progress
(Measurement Busy is low).
• A new sweep start trigger occurred while a sweep operation was in progress (Sweep
Busy is low).
If sampling errors occur, the trigger generation timing may be too fast. If the sampling
error indicator illuminates, check the selected trigger source, timer period, and so on.
2.6 Triggering
2-28
IM 765601-01E