16-26
IM 765601-01E
[:CHANnel<n>]:SENSe:FUNCtion/?
Function Sets the measurement function (voltage or
current) or queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:FUNCtion VOLTage|
CURRent
VOLTage
Selects voltage.
CURRent
Selects current.
[:CHANnel<n>]:SENSe:FUNCtion?
→ VOLT
Set to voltage.
→ CURR
Set to current.
Example
:SENS:FUNC VOLT
:CHAN2:SENS:FUNC CURR
:CHAN1:SENS:FUNC?
[:CHANnel<n>]:SENSe:[:VOLTage]:
RANGe/?
Function Sets the voltage measurement range or queries
the current setting.
Syntax
[:CHANnel<n>]:SENSe:[:VOLTage]:RANGe
<voltage>|MINimum|MAXimum|UP|DOWN
<voltage> Specify any voltage range.
The smallest range that includes the
specified voltage will be selected.
MINimum Sets the minimum value.
MAXimum Sets the maximum value.
UP
Increases the range by one level.
DOWN
Decreases the range by one level.
[:CHANnel<n>]:SENSe:[:VOLTage]:RANGe?
→ 200E-3 Set to 200 mV range.
→ 2E+0
Set to 2 V range.
→ 7E+0
Set to 7 V range (765601/765602).
→ 18E+0
Set to 18 V range (765601/765602).
→ 20E+0
Set to 20 V range (765611/765612).
→ 50E+0
Set to 50 V range (765611/765612).
Example
:SENS:VOLT:RANG 7
:CHAN2:SENS:VOLT:RANG MAX
:CHAN1:SENS:VOLT:RANG 200mV
:CHAN1:SENS:VOLT:RANG?
Description If VOLTage is omitted, the GS820 assumes that
the present measurement function was specified.
[:CHANnel<n>]:SENSe:[:VOLTage]:
RANGe:AUTO/?
Function Turns ON/OFF the auto measurement range or
queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:[:VOLTage]:RANGe:
AUTO 1|0|ON|OFF
1 or ON
Enables auto range.
0 or OFF
Enables fixed range.
[:CHANnel<n>]:SENSe:[:VOLTage]:RANGe:
AUTO?
→ 1 Set to auto range.
→ 0 Set to fixed range.
Example
:SENS:RANG:AUTO ON
:CHAN2:SENS:VOLT:RANG:AUTO 0
:CHAN1:SENS:VOLT:RANG:AUTO?
[:CHANnel<n>]:SENSe:[:CURRent]:
RANGe/?
Function Sets the current measurement range or queries
the current setting.
Syntax
[:CHANnel<n>]:SENSe:[:CURRent]:RANGe
<current>|MINimum|MAXimum|UP|DOWN
<current> Specify any current range.
The smallest range that includes the
specified current will be selected.
MINimum Sets the minimum value.
MAXimum Sets the maximum value.
UP
Increases the range by one level.
DOWN
Decreases the range by one level.
[:CHANnel<n>]:SENSe:[:CURRent]:RANGe?
→ 200E-9 Set to 200 nA range.
→ 2E-6
Set to 2 μA range.
→ 20E-6
Set to 20 μA range.
→ 200E-6 Set to 200 μA range.
→ 2E-3
Set to 2 mA range.
→ 20E-3
Set to 20 mA range.
→ 200E-3 Set to 200 mA range.
→ 600E-3 Set to 0.5 A range (765611/765612).
→ 1.2E+0 Set to 1 A range.
→ 3.2E+0 Set to 3 A range (765601/765602).
Example
:SENS:CURR:RANG 1.2
:CHAN2:SENS:CURR:RANG MAX
:CHAN2:SENS:CURR:RANG 20mA
:CHAN2:SENS:CURR:RANG?
Description If CURRent is omitted, the GS820 assumes that
the present measurement function was specified.