![YOKOGAWA GS820 User Manual Download Page 231](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275231.webp)
16-38
IM 765601-01E
:SYSTem:CLOCk:TIME/?
Function Sets the time or queries the current setting.
Syntax
:SYSTem:CLOCk:TIME <character string>
<character string> Specify a character string in
“hh:mm:ss” format.
hh = Hours (00 to 23)
mm = Minutes (00 to 59)
ss = Seconds (00 to 59)
:SYSTem:CLOCk:TIME?
→ <character string>
Time “hh:mm:ss” format
Example
:SYST:CLOC:TIME "17:30:45"
:SYST:CLOC:TIME?
:SYSTem:CLOCk:ADJust
Function Executes the ±30-s correction.
Syntax
:SYSTem:CLOCk:ADJust
Example
:SYST:CLOC:ADJ
:SYSTem:SETup:SAVE
Function Saves the setup data.
Syntax
:SYSTem:SETup:SAVE <character string>
<character string> The name of the setup data
file to be saved.
Example
:SYST:SET:SAVE "XY_Test.txt"
:SYSTem:SETup:LOAD
Function Loads the setup data.
Syntax
:SYSTem:SETup:LOAD <character string>
<character string> The name of the setup data
file to be loaded.
Example
:SYST:SET:LOAD "XY_Test.txt"
:SYSTem:SETup:CATalog?
Function Queries the list of setup data files.
Syntax
:SYSTem:SETup:CATalog?
→ <character string>,...
A list of setup data file names.
Example
:SYST:SET:CAT?
:SYSTem:SETup:DELete
Function Deletes the setup data file.
Syntax
:SYSTem:SETup:DELete <character
string>
<character string> The name of the setup data
file to be deleted.
Example
:SYST:SET:DEL "XY_Test.txt"
:SYSTem:SETup:PON/?
Function Sets the setup at power-on or queries the current
setting.
Syntax
:SYSTem:SETup:PON <character string>
<character string> Name of the setup data file
that is used to start up the
GS820.
:SYSTem:SETup:PON?
Example
:SYST:SET:PON "XY_Test.txt"
:SYSTem:ERRor?
Function Queries the error code and description.
Syntax
:SYSTem:ERRor?
→ <integer>,<character string>
Error code and error message.
Example
:SYST:ERR?
:SYSTem:LOCal
Function Switches to local mode.
Syntax
:SYSTem:LOCal
Example
:SYST:LOC
:SYSTem:REMote
Function Switches to remote mode.
Syntax
:SYSTem:REMote
Example
:SYST:REM
:SYSTem:KLOCk/?
Function Turns ON/OFF the key lock function or queries the
current setting.
Syntax
:SYSTem:KLOCk 1|0|ON|OFF
1 or ON
Enables key lock.
0 or OFF
Releases the key lock.
:SYSTem:KLOCk?
→ 1 Key lock enabled.
→ 0 Key lock disabled.
Example
:SYST:KLOC ON
:SYST:KLOC?
16.2 Commands