2.9
Store/Recall (Statistical Computation Value
Display)
Executing and Stopping the Storage Operation
(See section 9.1 for the procedure)
This function stores the measured results in the storage memory from when the storage
is turned ON until the specified number of points are stored. The storage count can be
set in the range of 1 to 100000. If the storage operation is aborted before reaching the
specified storage count, the stored results up to that point is stored. If the stored results
file generation function is turned ON, the stored results are output to a result file (Result.
csv) on the built-in GS820RAM (see the description of the USB storage function on page
2-36) when the storage is completed. If the stored results file generation function is OFF,
the stored results are not converted into a file. In this case, the stored results are read
using the communication function.
Result File
(See section 9.1 for the procedure)
If the stored results file generation function is turned ON before starting the storage
operation, the contents of the storage memory are output to a result file (Result.csv) on
the built-in GS820RAM when the storage is completed. A result file includes the elapsed
times from the storage start point (timestamps), digital output values, digital input values,
source function, source level, measurement function, measured values, comparison
lower limit, comparison upper limit, and comparison results.
Example of a Result File
Timestamp
Source function
Source level
Measurement function
Measured value
Comparison operation result
1
1 Displays the results only when comparison operation is ON.
Lower limit for comparison
1
Upper limit for comparison
1
Digital input value
Digital output value
CH1
CH2
Note
• Only the stored result of channels on which measurement is executed are output to the
result file. If the storage count of CH1 and CH2 are both zero, a result file is not created.
• If the measurement function is set to Auto, the measurement function that is indicated in the
result file is as follows:
If the source function is voltage (V), the measurement function is indicated as current (I).
If the source function is current (I), the measurement function is indicated as voltage (V).
• If the measurement function is set to Auto and a limiter is activated, the measurement
function switches to the source function. The source function when a limiter is activated is
indicated as follows:
If the source function is voltage (V), the source function is indicated as current (I).
If the source function is current (I), the source function is indicated as voltage (V).
The limit level is indicated for the source level when a limiter is activated.
2-33
IM 765601-01E
Explanation of Functions