7.5
Setting the Measurement Delay
Procedure
1.
Press
CONFIG
in the MEASURE section of the front panel to display the CONFIG
menu.
2.
Press the
M.Delay
soft key to display the measurement delay setup screen.
3.
Use the
rotary knob
or
numeric keys&< >
to set the measurement delay.
If you use the numeric keys, press the soft key for the desired unit to apply the
setting.
Explanation
Measurement Delay
A wait time from when the measurement trigger is generated until the measurement is
actually started. Set this delay if you want to insert a wait time after the source level is
changed until the measurement is actually started to allow the DUT to stabilize.
Computation
Measurement
delay
Measurement trigger
Time
Measurement
Selectable Range
0 μs to 3600.000000 s
Note
The time from when the source trigger is applied to when the source level stabilizes varies
depending on the load, source range, and limiter level. Keep these in mind in addition to the
time needed for the DUT to stabilize after the source level is applied to adjust the length of the
measurement delay.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SENSe:DELay <time>|MINimum|MAXimum
7-6
IM 765601-01E