Linear Sweep
(See section 6.1 for the procedure)
Performs a linear sweep operation at step levels at regular intervals from the specified
start level to the stop level.
Step level
Stop level
Start level
When source waveform is DC
Trigger count
Source level
Source level
Sweep start
Trigger count
Sweep start
When source waveform is pulse
Stop level
Start level
Step level
Pulse base value
Trigger Trigger Trigger Trigger Trigger
Trigger Trigger Trigger Trigger Trigger
Start Level, Stop Level, and Step Level
• Setting Resolution of the Start level, Stop Level, and Step level for Voltage
Source Mode
765601/765602
|Start Level|, |Stop Level|, and Step Level
Setting Resolution
0.000 mV ≤ X ≤ 200.000 mV
1 µV
0.20000 V < X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 7.0000 V
100 µV
7.0000 V
< X ≤ 18.0000 V
100 µV
765611/765612
|Start Level|, |Stop Level|, and Step Level
Setting Resolution
0.000 mV ≤ X ≤ 200.000 mV
1 µV
0.20000 V < X ≤ 2.00000 V
10 µV
2.0000 V
< X ≤ 20.0000 V
100 µV
20.0000 V < X ≤ 50.0000 V
100 µV
• Setting Resolution of the Start level, Stop Level, and Step level for Current
Source Mode
765601/765602
|Start Level|, |Stop Level|, and Step Level
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 3.20000 A
10 µA
765611/765612
|Start Level|, |Stop Level|, and Step Level
Setting Resolution
0.000 nA
≤ X ≤ 200.000 nA
1 pA
0.20000 µA < X ≤ 2.00000 µA
10 pA
2.0000 µA
< X ≤ 20.0000 µA
100 pA
20.000 µA
< X ≤ 200.000 µA
1 nA
0.20000 mA < X ≤ 2.00000 mA
10 nA
2.0000 mA < X ≤ 20.0000 mA
100 nA
20.000 mA < X ≤ 200.000 mA
1 µA
0.20000 A
< X ≤ 1.20000 A
10 µA
2.5 Sweep
2-20