6.5
Starting the Sweep Operation
Procedure
The sweep operation starts when a sweep start source is applied or when you press
START
.
Explanation
Starting the Liner Sweep, Log Sweep, or Programmable Sweep
The initial state is Waiting for Sweep Start. The GS820 will be able to receive the source
trigger once the sweep operation is started from the Waiting for Sweep Start state. After
receiving the source trigger the number of times equal to sweep count × repeat count,
the sweep operation stops, and the GS820 returns to the Waiting for Sweep Start state.
Note
• Source triggers that are received during the Waiting for Sweep Start state are discarded.
• If a new sweep start occurs while the sweep operation is in progress, the on-going sweep
operation is aborted, and an sampling error occurs.
<<Corresponding Command Mnemonic>>
:STARt
6-12