![YOKOGAWA GS820 User Manual Download Page 38](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275038.webp)
Current Range
The current range consists of the current source range and current limiter range.
V
I
3.2 A/7 V range
1.2 A/18 V range
200 mA/18 V range
20 mA/18 V range
2 mA/18 V range
200 µA/18 V range
20 µA/18 V range
2 µA/18 V range
200 nA/18 V range
Range Source Range Resolution Max. Load Voltage
200 nA ±200.000 nA
1 pA
±18 V
2 µA
±2.00000 µA
10 pA
±18 V
20 µA
±20.0000 µA
100 pA
±18 V
200 µA ±200.000 µA
1 nA
±18 V
2 mA
±2.00000 mA
10 nA
±18 V
20 mA
±20.0000 mA
100 nA
±18 V
200 mA ±200.000 mA
1 µA
±18 V
1 A
±1.20000 A
10 µA
±18 V
3 A
±3.20000 A
10 µA
±7 V
765601/765602
V
I
0.5 A/50 V range
200 mA/50 V range
20 mA/50 V range
2 mA/50 V range
200 µA/50 V range
20 µA/50 V range
2 µA/50 V range
200 nA/50 V range
Range Source Range Resolution Max. Load Voltage
200 nA ±200.000 nA
1 pA
±50 V
2 µA
±2.00000 µA
10 pA
±50 V
20 µA
±20.0000 µA
100 pA
±50 V
200 µA ±200.000 µA
1 nA
±50 V
2 mA
±2.00000 mA
10 nA
±50 V
20 mA
±20.0000 mA
100 nA
±50 V
200 mA ±200.000 mA
1 µA
±50 V
0.5 A
±0.60000 A
10 µA
±50 V
1.0 A
±1.20000 A
10 µA
±20 V
765611/765612
1.2 A/20 V range
Source Function
(See section 5.1 for operating procedure)
The source function is voltage (VS) or current (IS).
Voltage (VS): Operates as a constant voltage source. The current limiter is enabled.
Current (IS): Operates as a constant current source. The voltage limiter is enabled.
2.3 Source
2-8
IM 765601-01E