![YOKOGAWA GS820 User Manual Download Page 213](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275213.webp)
16-20
IM 765601-01E
[:CHANnel<n>]:SOURce[:VOLTage]:
PULSe:BASE/?
Function Sets the pulse base value for voltage pulse source
mode or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:PULSe:
BASE <voltage>|MINimum|MAXimum
<voltage> Specify any voltage pulse base
value.
MINimum Sets the minimum value.
In fixed range mode, it is set to the
minimum value within the range.
In auto range mode, it is set to
‒18 V (765601/765602) or ‒50 V
(765611/765612).
MAXimum Sets the maximum value.
In fixed range mode, it is set to the
maximum value within the range.
In auto range mode, it is set to
18 V (765601/765602) or 50 V
(765611/765612).
[:CHANnel<n>]:SOURce[:VOLTage]:PULSe:
BASE?
→ <voltage> The present voltage pulse base
value.
Example
:SOUR:VOLT:PULS:BASE -1.0E+2
:CHAN2:SOUR:VOLT:PULS:BASE -250mV
:CHAN1:SOUR:VOLT:PULS:BASE?
Description If VOLTage is omitted, the GS820 assumes that
the present source function was specified.
[:CHANnel<n>]:SOURce[:VOLTage]:ZERO:
IMPedance/?
Function Sets the impedance (high or low) for voltage zero
source or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:ZERO:
IMPedance HIGH|LOW
HIGH Sets to high impedance.
LOW Sets to low impedance.
[:CHANnel<n>]:SOURce[:VOLTage]:ZERO:
IMPedance?
→ HIGH The voltage zero source is set to high
impedance.
→ LOW The voltage zero source is set to low
impedance.
Example
:SOUR:VOLT:ZERO:IMP HIGH
:CHAN2:SOUR:VOLT:ZERO:IMP LOW
:CHAN1:SOUR:VOLT:ZERO:IMP?
Description If VOLTage is omitted, the GS820 assumes that
the present source function was specified.
16.2 Commands
[:CHANnel<n>]:SOURce[:CURRent]:
RANGe/?
Function Sets the current source range or queries the
current setting.
Syntax
[:CHANnel<n>]:SOURce[:CURRent]:RANGe
<current>|MINimum|MAXimum|UP|DOWN
<current> Specify any current range.
The smallest range that includes the
specified current will be selected.
MINimum Sets the minimum value.
MAXimum Sets the maximum value.
UP
Increases the range by one level.
DOWN
Decreases the range by one level.
[:CHANnel<n>]:SOURce[:CURRent]:RANGe?
→ 200E-9 Set to 200 nA range.
→ 2E-6
Set to 2 μA range.
→ 20E-6
Set to 20 μA range.
→ 200E-6 Set to 200 μA range.
→ 2E-3
Set to 2 mA range.
→ 20E-3
Set to 20 mA range.
→ 200E-3 Set to 200 mA range.
→ 600E-3 Set to 0.5 A range (765611/765612).
→ 1.2E+0 Set to 1 A range.
→ 3.2E+0 Set to 3 A range (765601/765602).
Example
:SOUR:CURR:RANG 200E-3
:CHAN2:SOUR:CURR:RANG MAX
:CHAN1:SOUR:CURR:RANG 20uA
:CHAN1:SOUR:CURR:RANG?
Description If CURRent is omitted, the GS820 assumes that
the present source function was specified.
[:CHANnel<n>]:SOURce[:CURRent]:
RANGe:AUTO/?
Function Turns ON/OFF the auto current source range or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:CURRent]:RANGe:
AUTO 1|0|ON|OFF
1 or ON
Enables auto range.
0 or OFF
Enables fixed range.
[:CHANnel<n>]:SOURce[:CURRent]:RANGe:
AUTO?
→ 1 Set to auto range.
→ 0 Set to fixed range.
Example
:SOUR:CURR:RANG:AUTO ON
:CHAN2:SOUR:CURR:RANG:AUTO 0
:CHAN1:SOUR:CURR:RANG:AUTO?
Description If CURRent is omitted, the GS820 assumes that
the present source function was specified.