Specifications
Built-in Variables
<Variable> Meaning
<Reference> <Substitution>
TM
A real number that represents the elapsed seconds from
00:00:00 at January 1, 1970 in 1-µs resolution.
Yes
No
SF
Source function
0: voltage, 1: current
Yes
Yes (setting changed immediately)
SR
Source range
0: 200-mV range to 3: 18-V range
1
Yes
Yes (setting changed immediately)
0: 200-nA range to 8: 3-A range
1
SL
Source level
Yes
Yes (setting changed immediately)
SD
Source delay
Yes
Yes (setting changed immediately)
PW
Pulse width
Yes
Yes (setting changed immediately)
PB
Pulse base
Yes
Yes (setting changed immediately)
MF
Measurement function
0: voltage, 1: current
Yes
Yes (setting changed immediately)
MR
Measurement range
0: 200-mV range to 3: 18-V range
1
Yes
Yes (setting changed immediately)
0: 200-nA range to 8: 3-A range
1
ML
Measured value
Yes
Yes (manipulates the measurement result)
MD
Measurement delay
Yes
Yes (setting changed immediately)
HL
High limiter
Yes
Yes (setting changed immediately)
LL
Low limiter
Yes
Yes (setting changed immediately)
LS
Limiter status
−1: Low, 0: Limiter not activated, 1: High
Yes
No
OS
Output state
0: OFF, 1: ON, 2: Zero
Yes
Yes (setting changed immediately)
T1
Timer 1
Yes
Yes (setting changed immediately)
T2
Timer 2
Yes
Yes (setting changed immediately)
DO
Digital output
0 to 65535
Yes
Yes (setting changed immediately)
DI
Digital input
0 to 65535
Yes
No
AT
Auxiliary trigger generation
No
Yes
(generates an auxiliary trigger immediately)
V
Source level or measured value in voltage dimension
Yes
No
I
Source level or measured value in current dimension
Yes
No
A to C
Constant. Allows a value to be entered from the panel when
used in a statement.
Yes
Yes
J
A counter. Initial value is zero. Incremented each time a
measurement is completed.
Yes
No
1 Fot the 765611/765612, 0: 200-mV range to 3: 50-V range
0: 200-nA range to 8: 1-A range
Built-in Constants
E
The base of the natural logarithm 2.7182818...
PI
Ratio of circumference to diameter 3.1415926...
NAN Not A Number. Indicates “not measured” if substituted into ML.
INF Infinity. “OVER” display occurs if substituted into ML.
Variables
D, F to H, K to U, and W to Z
Binary Operators
+
Returns the sum of the left and right expressions.
− Returns the difference between the left and right expressions.
*
Returns the product of the left and right expressions.
/ Returns the quotient of the left and right expressions.
% Returns the remainder of the quotient of the left and right expressions.
^
Returns the left term to the power of the right expressions.
| Returns the logical OR of the left and right expressions.
& Returns the logical AND of the left and right expressions.
Unary Operators
− Returns the negation of the right expression.
~
Returns the bit inversion of the left expression.
!
Returns the logical inversion of the left expression.
8.5 User-Defined Computation
8-10
IM 765601-01E