6.
Use the
rotary knob
or
numeric keys&< >
to set the parameter.
7.
If you use the numeric keys, press the
Enter
soft key to confirm the setting.
Viewing the Computation Definition
8.
Press the
View
soft key to display the contents of the selected computation
definition.
Explanation
The measured results can be used to perform various computations. The computation
definition file can be any of the built-in computation files or a user-defined computation
file that you create. For a description of how to write a user-defined computation file, see
section 8.5, “User-Defined Computation.”
Built-in Computation Files
The following five equations are built in.
• Decibel (dB.txt)
ML = A*log(ML/B): Computes the decibel value of the measured value with respect to B.
• Percent (Percent.txt)
ML = (ML/A)*100: Computes the percentage of the measured value with respect to A.
• Scaling (Scaling.txt)
ML = A*ML+B:
Scales the measured value.
• Power (Power.txt)
ML = V*I: Calculates the power. However, the power cannot be calculated correctly if
the source function and measurement function are the same or when the
GS820 is in voltmeter, ammeter, or resistance meter mode. Set the GS820
to voltage source and current measurement modes or current source and
voltage measurement modes.
• Resistance (Resistance.txt)
ML = V/I: Calculates the resistance. However, the resistance cannot be calculated
correctly if the source function and measurement function are the same or
when the GS820 is in voltmeter, ammeter, or resistance meter mode. Set
the GS820 to voltage source and current measurement modes or current
source and voltage measurement modes.
Selectable Range of Parameters
−9.24 to +9.24
Viewing the Computation Definition
The contents of the computation definition files stored in the MATH directory of
GS820ROM can be viewed.
8.3 Equation Computation
8-4
IM 765601-01E