5.10 Setting the Pulse Base
Procedure
1.
Press
SHIFT+CONFIG (PULSE)
in the SOURCE section of the front panel to
display the PULSE menu.
2.
Press the
Pulse Base
soft key to display the pulse base setup screen.
3.
Use the
rotary knob
or
numeric keys&< >
to set the pulse base value.
If you use the numeric keys, press the soft key for the desired unit to confirm the
setting.
Explanation
Selectable Range of Pulse Base Value
See the source range for fixed range in section 5.2, “Setting the Source Range.”
Note
If the source range is set to auto range, the range appropriate for the larger of the two values
|source value| and |pulse base value| is applied.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SOURce[:VOLTage]:PULSe:BASE <voltage>|MINimum|MAXimum
[:CHANnel<n>]:SOURce[:CURRent]:PULSe:BASE <current>|MINimum|MAXimum
5-16