3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
2.4 Measurement
Measurement Function and Measurement Range
Measurement Function
(See section 7.2 for operating procedure)
The measurement function is voltage (VM) or current (IM). Switching the measurement
function does not affect the source.
Voltage (VM): Measures voltage.
Current (IM): Measures current.
Measurement Range
(See section 7.3 for the procedure)
When Measuring Voltage
765601/765602
765611/765612
Range
Measurement Range Resolution
Range
Measurement Range Resolution
200 mV
±210.000 mV
1 µV
200 mV
±210.000 mV
1 µV
2 V
±2.100000 V
10 µV
2 V
±2.100000 V
10 µV
7 V
±7.1000 V
100 µV
20 V
±21.0000 V
100 µV
18 V
±18.0000 V
100 µV
50 V
±50.1000 V
100 µV
When Measuring Current
765601/765602
765611/765612
Range
Measurement Range Resolution
Range
Measurement Range Resolution
200 nA
±210.000 nA
1 pA
200 nA
±210.000 nA
1 pA
2 µA
±2.10000 µA
10 pA
2 µA
±2.10000 µA
10 pA
20 µA
±21.0000 µA
100 pA
20 µA
±21.0000 µA
100 pA
200 µA
±210.000 µA
1 nA
200 µA
±210.000 µA
1 nA
2 mA
±2.10000 mA
10 nA
2 mA
±2.10000 mA
10 nA
20 mA
±21.0000 mA
100 nA
20 mA
±21.0000 mA
100 nA
200 mA
±210.000 mA
1 µA
200 mA
±210.000 mA
1 µA
1 A
±1.30000 A
10 µA
0.5 A
±0.60000 A
10 µA
3 A
±3.20000 A
10 µA
1.0 A
±1.20000 A
10 µA
Fixed Measurement Range and Auto Range
There are two measurement ranges: fixed range and auto range.
Fixed Measurement Range
The target range is specified directly. This setting is used when you know the range of
the measured value in advance. The measurable level is within the measurement range
of the specified range (see “Measurement Range” described earlier). However, if the
measurement function is the same as the source function, the measurement range is
fixed to the source range. If the input is outside the present measurement range, over
range results.
Auto Measurement Range
The GS820 automatically switches to the most suitable range according to the measured
value. This setting is used when you do not know the range of the measured value. This
setting allows you to make measurements without having to worry about the range.
However, measurement takes longer, because additional measurements to detect the
range are performed.
2-15
IM 765601-01E
Explanation of Functions