16-16
IM 765601-01E
[:CHANnel<n>]:SOURce:MODE/?
Function Sets the source mode (constant, sweep,
programmable, or single-step) or queries the
current setting.
Syntax
[:CHANnel<n>]:SOURce:MODE
FIXed|SWEep|LIST|SINGle
FIXed
Selects fixed value (sweep off).
SWEep
Selects linear or log sweep.
LIST
Selects programmable sweep.
SINGle
Selects single-step sweep.
[:CHANnel<n>]:SOURce:MODE?
→ FIX
Set to fixed value (sweep off).
→ SWE
Set to linear or log sweep.
→ LIST
Set to programmable sweep.
→ SING
Set to single-step sweep.
Example
:SOUR:MODE LIST
:CHAN2:SOUR:MODE FIX
:CHAN1:SOUR:MODE?
[:CHANnel<n>]:SOURce:DELay/?
Function Sets the source delay or queries the current
setting.
Syntax
[:CHANnel<n>]:SOURce:DELay
<time>|MINimum|MAXimum
<time>
Specify any source delay value.
MINimum Sets the minimum value (= 15 μs).
MAXimum Sets the maximum value (= 3600 s).
[:CHANnel<n>]:SOURce:DELay?
→ <time> The present source delay time
Example
:SOUR:DEL 2.5E-3
:CHAN2:SOUR:DEL MIN
:CHAN1:SOUR:DEL 1.25ms
:CHAN1:SOUR:DEL?
[:CHANnel<n>]:SOURce:LIST:SELect/?
Function Sets the programmable sweep pattern file or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce:LIST:SELect
<character string>|NONE
<character string> Specify a pattern file name.
NONE
Pattern file not selected condition.
[:CHANnel<n>]:SOURce:LIST:SELect?
→ <character string>
The present pattern file name
→ NONE
Pattern file not selected.
Example
:SOUR:LIST:SEL "Test1.csv"
:CHAN2:SOUR:LIST:SEL "Test2.csv"
:CHAN1:SOUR:LIST:SEL?
[:CHANnel<n>]:SOURce:LIST:CATalog?
Function Queries the list of programmable sweep pattern
files.
Syntax
[:CHANnel<n>]:SOURce:LIST:CATalog?
→ <character string>,... Pattern file name list
Example
:SOUR:LIST:CAT?
[:CHANnel<n>]:SOURce:LIST:DELete
Function Deletes a programmable sweep pattern file.
Syntax
[:CHANnel<n>]:SOURce:LIST:DELete
<character string>
<character string> The name of the pattern file
to be deleted.
Example
:SOUR:LIST:DEL "Test1.csv"
[:CHANnel<n>]:SOURce:LIST:LOAD
Function Loads the pattern data of a programmable sweep.
Syntax
[:CHANnel<n>]:SOURce:LIST:LOAD
<character string>|<block data>
<character string> Pattern data character string
to be loaded.
<block data> Block data expression of the
pattern data to be loaded.
Example
:SOUR:LIST:LOAD #40062...
:CHAN2:SOUR:LIST:LOAD #500128...
[:CHANnel<n>]:SOURce[:VOLTage]:
RANGe/?
Function Sets the voltage source range or queries the
current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:RANGe
<voltage>|MINimum|MAXimum|UP|DOWN
<voltage> Specify any voltage range.
The smallest range that includes the
specified voltage will be selected.
MINimum Sets the minimum value.
MAXimum Sets the maximum value.
UP
Increases the range by one level.
DOWN
Decreases the range by one level.
[:CHANnel<n>]:SOURce[:VOLTage]:RANGe?
→ 200E-3 Set to 200 mV range.
→ 2E+0
Set to 2 V range.
→ 7E+0
Set to 7 V range (765601/765602).
→ 18E+0
Set to 18 V range (765601/765602).
→ 20E+0
Set to 20 V range (765611/765612).
→ 50E+0
Set to 50 V range (765611/765612).
Example
:SOUR:VOLT:RANG 2E+0
:CHAN2:SOUR:VOLT:RANG 7V
:CHAN1:SOUR:VOLT:RANG MAX
:CHAn1:SOUR:VOLT:RANG?
Description If VOLTage is omitted, the GS820 assumes that
the present source function was specified.
16.2 Commands