![YOKOGAWA GS820 User Manual Download Page 106](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275106.webp)
For Pulse Source Mode
The GS820 outputs the pulse base value in steady-state condition and outputs the
source level of a specified pulse width after the source delay elapses from the trigger
point.
Time
Source
Pulse
base value
Source level
Pulse width
Source
delay
↑
Source
trigger
For Linear Sweep, Log Sweep, and Programmable Sweep
The GS820 executes a step within a predefined pattern sweep after the source delay
elapses from the source trigger point.
Source
delay
Source
delay
↑
Source
trigger
Source
Time
↑
Source
trigger
For Single-Step Sweep
After the setting is changed from the front panel or communication command, the setting
change is actually executed after the source delay elapses from the source trigger point.
Source
delay
Source
Time
Next source level
Previous
source
level
↑
Setting
change
↑
Source
trigger
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SOURce:DELay <time>|MINimum|MAXimum
5.5 Setting the Source Delay
5-11
IM 765601-01E