16-25
IM 765601-01E
Communication Commands
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
16.2.5 Measurement Commands (SENSe Group)
[:CHANnel<n>]:SENSe:TRIGger/?
Function Sets the measurement trigger source (source
change, sweep end, auxiliary trigger, Timer1, 2, or
immediate) or queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:TRIGger SOURce|
SWEep|AUXiliary|TIMer1|TIMer2|
IMMediate
SOURce
Selects source change.
SWEep
Selects sweep end.
AUXiliary
Selects auxiliary trigger.
TIMer1
Selects Timer1 (constant period).
TIMer2
Selects Timer2 (constant period).
IMMediate Selects immediate trigger.
[:CHANnel<n>]:SENSe:TRIGger?
→ SOUR
Set to source change.
→ SWE
Set to sweep end.
→ AUX
Set to auxiliary trigger.
→ TIM1
Set to Timer1.
→ TIM2
Set to Timer2.
→ IMM
Set to immediate trigger.
Example
:SENS:TRIG AUX
:CHAN2:SENS:TRIG TIM2
:CHAn1:SENS:TRIG?
[:CHANnel<n>]:SENSe:TRIGger:
AUXiliary:POLarity/?
Function Sets the auxiliary trigger polarity (normal or
inverted) or queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:TRIGger:
AUXiliary:POLarity NORMal|INVerted
NORMal
Selects normal (falling edge).
INVerted
Selects inverted (rising edge).
[:CHANnel<n>]:SENSe:TRIGger:
AUXiliary:POLarity?
→ NORM Set to normal (falling edge).
→ INV
Set to inverted (rising edge).
Example
:SENS:TRIG:AUX:POL INV
:CHAN2:SENS:TRIG:AUX:POL NORM
:CHAN1:SENS:TRIG:AUX:POL?
[:CHANnel<n>]:SENSe[:STATe]/?
Function Turns ON/OFF the measurement function or
queries the current setting.
Syntax
[:CHANnel<n>]:SENSe[:STATe] 1|0|ON|
OFF
1 or ON
Turns the measurement function ON.
0 or OFF
Turns the measurement function
OFF.
[:CHANnel<n>]:SENSe[:STATe]?
→ 1 The measurement function is ON.
→ 0 The measurement function is OFF.
Example
:SENS ON
:CHAN2:SENS:STAT 0
:CHAN1:SENS:STAT?
[:CHANnel<n>]:SENSe:MODE/?
Function Sets the measurement mode (fixed function, auto
function, voltmeter, ammeter, or resistance meter)
or queries the current setting.
Syntax
[:CHANnel<n>]:SENSe:MODE FIXed|AUTO|
VMETer|IMETer|RMETer
FIXed
Selects the fixed function mode.
AUTO
Selects the auto function mode.
VMETer
Selects the voltmeter mode.
IMETer
Selects the ammeter mode.
RMETer
Selects the resistance meter mode.
[:CHANnel<n>]:SENSe:MODE?
→ FIX
Set to fixed function.
→ AUTO
Set to auto function.
→ VMET
Set to voltmeter mode.
→ IMET
Set to ammeter mode.
→ RMET
Set to resistance meter mode.
Example
:SENS:MODE AUTO
:CHAN2:SENS:MODE RMET
:CHAN1:SENS:MODE?
16.2 Commands