5.2
Setting the Source Range
Procedure
There are two source ranges: fixed range and auto range.
Changing the Fixed Range
1.
Press the
or
key under
RANGE
in the SOURCE section of the front panel to
change the source range.
Turning Auto Range ON/OFF
1.
Press
SHIFT+
(AUTO)
in the SOURCE section of the front panel.
The auto range function turns ON and Auto appears in the display.
If you press
SHIFT+
(AUTO)
again or if you press
or
to change the range when
auto range is ON, auto range turns OFF and fixed range is enabled.
Explanation
Display Example of Source Range
Fixed range display
Display when auto range is ON
Fixed Range
765601/765602
•
Voltage Source Range
Source Range
Range Generated
Resolution
Max. Load Current
200 mV
±200.000 mV
1 µV
±3.2 A
2 V
±2.00000 V
10 µV
±3.2 A
7 V
±7.0000 V
100 µV
±3.2 A
18 V
±18.0000 V
100 µV
±1.2 A
•
Current Source Range
Source Range
Range Generated
Resolution
Max. Load Voltage
200 nA
±200.000 nA
1 pA
±18 V
2 µA
±2.00000 µA
10 pA
±18 V
20 µA
±20.0000 µA
100 pA
±18 V
200 µA
±200.000 µA
1 nA
±18 V
2 mA
±2.00000 mA
10 nA
±18 V
20 mA
±20.0000 mA
100 nA
±18 V
200 mA
±200.000 mA
1 µA
±18 V
1A
±1.20000 A
10 µA
±18 V
3A
±3.20000 A
10 µA
±7 V
5-2