7.6
Selecting the Measurement Trigger
Procedure
1.
Press
CONFIG
in the MEASURE section of the front panel to display the CONFIG
menu.
2.
Press the
MeasTrig
soft key to display the Measure Trigger menu.
Press the
Next 1/2
soft key to see the next page of available choices.
3.
Press the soft key corresponding to the desired measurement trigger.
Explanation
The measurement trigger is a signal source used to start the measurement. Select from
the items below.
SrcChg: Source change
Aux
:
Auxiliary trigger (rising edge)
Aux
:
Auxiliary trigger (falling edge)
Timer1
Timer2
Imm:
Immediate
SwpEnd: At sweep end
Note
A sampling error occurs if a new measurement trigger is generated while a measurement is in
progress.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SENSe:TRIGger SOURce|SWEep|AUXiliary|TIMer1|TIMer2|IMMediate
[:CHANnel<n>]:SENSe:TRIGger:AUXiliary:POLarity NORMal|INVerted
7-7