18.1 Source Section
DC Voltage Source
765601/765602
Range Range Generated
Resolution
Max. Load Current
Accuracy (One Year)
Temperature Coefficient
± (% of s V)
± (% of s V)/°C
200 mV ±200.000 mV
1 µV
±3.2 A
0.02 + 250 µV
0.003 + 35 µV
2 V
±2.00000 V
10 µV
±3.2 A
0.02 + 400 µV
0.003 + 60 µV
7 V
±7.0000 V
100 µV
±3.2 A
0.02 + 2 mV
0.003 + 300 µV
18 V
±18.0000 V
100 µV
±1.2 A
0.02 + 2 mV
0.003 + 300 µV
Accuracy: One year accuracy for 23 ± 5°C.
Temperature coefficient: Add the temperature coefficient at 5 to 18°C and 28 to 40°C.
Output resistance (for four-wire system remote sensing)
• 200 mV, 2 V range: (Shunt resistance*/40000) Ω or less
• 7 V, 18 V range: (Shunt resistance/5000) Ω or less
* For details on the shunt resistance, see “DC Current Measurement” in section 18.2, “Measurement Section.”
765611/765612
Range Range Generated
Resolution
Max. Load Current
Accuracy (One Year)
Temperature Coefficient
± (% of s V)
± (% of s V)/°C
200 mV ±200.000 mV
1 µV
±1.2 A
0.02 + 250 µV
0.003 + 35 µV
2 V
±2.00000 V
10 µV
±1.2 A
0.02 + 400 µV
0.003 + 60 µV
20 V
±20.0000 V
100 µV
±1.2 A
0.02 + 8 mV
0.003 + 300 µV
50 V
±50.0000 V
100 µV
±0.6 A
0.02 + 20 mV
0.003 + 3 mV
Accuracy: One year accuracy for 23 ± 5°C.
Temperature coefficient: Add the temperature coefficient at 5 to 18°C and 28 to 40°C.
Output resistance (for four-wire system remote sensing)
• 200 mV, 2 V range: (Shunt resistance*/40000) Ω or less
• 20 V, 50 V range: (Shunt resistance/2000) Ω or less
* For details on the shunt resistance, see “DC Current Measurement” in section 18.2, “Measurement Section.”
DC Current Source
765601/765602
Range Range Generated
Resolution
Max. Load Voltage
Accuracy (One Year)
Temperature Coefficient
± (% of s A)
± (% of s A)/°C
200 nA ±200.000 nA
1 pA
±18 V
0.06 + 3 nA
500 pA
2 µA
±2.00000 µA
10 pA
±18 V
0.04 + 3 nA
500 pA
20 µA
±20.0000 µA
100 pA
±18 V
0.03 + 3 nA
0.0045 + 450 pA
200 µA ±200.000 µA
1 nA
±18 V
0.03 + 30 nA
0.0045 + 4.5 nA
2 mA
±2.00000 mA
10 nA
±18 V
0.03 + 250 nA
0.0045 + 37.5 nA
20 mA
±20.0000 mA
100 nA
±18 V
0.03 + 2.5 µA
0.0045 + 375 nA
200 mA ±200.000 mA
1 µA
±18 V
0.03 + 25 µA
0.0045 + 3.75 µA
1 A
±1.20000 A
10 µA
±18 V
0.05 + 900 µA
0.0075 + 135 µA
3 A
±3.20000 A
10 µA
±7 V
0.05 + 1.5 mA
0.0075 + 225 µA
Accuracy: One year accuracy for 23 ± 5°C.
Temperature coefficient: Add the temperature coefficient at 5 to 18°C and 28 to 40°C.
Output resistance
• 1 A, 3 A range: 10 kΩ or greater
• 20 μA to 200 mA range: (Shunt resistance* × 50000) Ω or greater
• 200 nA, 2 μA range: 10 GΩ or greater
* For details on the shunt resistance, see “DC Current Measurement” in section 18.2, “Measurement Section.”