16-17
IM 765601-01E
Communication Commands
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
App
Index
[:CHANnel<n>]:SOURce[:VOLTage]:
RANGe:AUTO/?
Function Turns ON/OFF the auto voltage source range or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:RANGe:
AUTO 1|0|ON|OFF
1 or ON
Enables auto range.
0 or OFF
Enables fixed range.
[:CHANnel<n>]:SOURce[:VOLTage]:RANGe:
AUTO?
→ 1 Set to auto range.
→ 0 Set to fixed range.
Example
:SOUR:VOLT:RANG:AUTO ON
:CHAN2:SOUR:VOLT:RANG:AUTO 0
:CHAN1:SOUR:VOLT:RANG:AUTO?
Description If VOLTage is omitted, the GS820 assumes that
the present source function was specified.
[:CHANnel<n>]:SOURce[:VOLTage]:
LEVel/?
Function Sets the voltage source level or queries the
current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:LEVel
<voltage>|MINimum|MAXimum
<voltage> Specify any voltage level.
MINimum Sets the minimum value.
In fixed range mode, it is set to the
minimum value within the range.
In auto range mode, it is set to
‒18 V (765601/765602) or ‒50 V
(765611/765612).
MAXimum Sets the maximum value.
In fixed range mode, it is set to the
maximum value within the range.
In auto range mode, it is set to
18 V (765601/765602) or 50 V
(765611/765612).
[:CHANnel<n>]:SOURce[:VOLTage]:LEVel?
→ <voltage> The present voltage level.
Example
:SOUR:VOLT:LEV 3.5
:CHAN2:SOUR:VOLT:LEV -1.5V
:CHAn1:SOUR:VOLT:LEV?
Description If VOLTage is omitted, the GS820 assumes that
the present source function was specified.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection[:STATe]/?
Function Turns ON/OFF the limiter function or queries the
current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection[:STATe] 1|0|ON|OFF
1 or ON
Turns the limiter function ON.
0 or OFF
Turns the limiter function OFF.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection[:STATe]?
→ 1 The limiter function is ON.
→ 0 The limiter function is OFF.
Example
:SOUR:VOLT:PROT ON
:CHAN2:SOUR:VOLT:PROT:STAT 0
:CHAN1:SOUR:VOLT:PROT:STAT?
Description If VOLTage is omitted, the GS820 assumes that
the present limit function was specified.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LINKage/?
Function Turns ON/OFF the limiter tracking function or
queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LINKage 1|0|ON|OFF
1 or ON
Turns the tracking function ON.
0 or OFF
Turns the tracking function OFF.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LINKage?
→ 1 The tracking function is ON.
→ 0 The tracking function is OFF.
Example
:SOUR:VOLT:PROT:LINK ON
:CHAN2:SOUR:VOLT:PROT:LINK 0
:CHAN1:SOUR:VOLT:PROT:LINK?
Description If VOLTage is omitted, the GS820 assumes that
the present limit function was specified.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LEVel/?
Function Sets the voltage limit value (for current source
mode) or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LEVel <voltage>|MINimum|
MAXimum
<voltage> Specify any level for the positive and
negative voltage limits.
MINimum Sets the minimum value.
MAXimum Sets the maximum value.
[:CHANnel<n>]:SOURce[:VOLTage]:
PROTection:LEVel?
→ <voltage> The present voltage limit.
Example
:SOUR:VOLT:PROT:LEV 2.0
:CHAN2:SOUR:VOLT:PROT:LEV 2.5V
:CHAN1:SOUR:VOLT:PROT:LEV?
Description If VOLTage is omitted, the GS820 assumes that
the present limit function was specified.
16.2 Commands