5.7
Selecting the Source Trigger
Procedure
1.
Press
CONFIG
in the SOURCE section of the front panel to display the CONFIG
menu.
2.
Press the
SrcTrig
soft key to display the Source Trigger menu.
3.
Press the soft key corresponding to the desired source trigger.
Explanation
Source Trigger
A signal source used to trigger the source operation. Select from the items below.
External: External trigger
Aux
:
Auxiliary trigger (rising edge)
Aux
:
Auxiliary trigger (falling edge)
Timer1
Timer2
MeasEnd: At the end of a measurement
Note
A sampling error occurs if a new source trigger is generated while a source action is in
progress.
<<Corresponding Command Mnemonic>>
[:CHANnel<n>]:SOURce:TRIGger EXTernal|AUXiliary|TIMer1|TIMer2|SENSe
[:CHANnel<n>]:SOURce:TRIGger:AUXiliary:POLarity NORMal|INVerted