5.9
Offset Calibration
Procedure
1.
Press
CONFIG
in the SOURCE section of the front panel to display the CONFIG
menu.
2.
Press the
ZeroCal Exec
soft key to execute the calibration.
The
ZeroCal Exec
indicator in the display is highlighted while the calibration is in progress.
Explanation
Offset calibration is executed to correct the offset drift in the source level that is caused
by temperature changes and the like.
Note
• Because measurements are performed to calibrate all the ranges in an offset calibration, the
source and measurement operations are suspended for a few seconds while the calibration
is in progress.
• The results of the calibration are lost when the power is turned OFF.
<<Corresponding Command Mnemonic>>
*CAL?
5-15
IM 765601-01E