![YOKOGAWA GS820 User Manual Download Page 217](http://html.mh-extra.com/html/yokogawa/gs820/gs820_user-manual_911275217.webp)
16-24
IM 765601-01E
[:CHANnel<n>]:SOURce[:CURRent]:
PULSe:BASE/?
Function Sets the pulse base value for current pulse source
mode or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:CURRent]:PULSe:
BASE <current>|MINimum|MAXimum
<current> Specify any current pulse base
value.
MINimum Sets the minimum value.
In fixed range mode, it is set to the
minimum value within the range.
In auto range mode, it is set to
‒3.2 A (765601/765602) or ‒1.2 A
(765611/765612).
MAXimum Sets the maximum value.
In fixed range mode, it is set to the
maximum value within the range.
In auto range mode, it is set to
3.2 A (765601/765602) or 1.2 A
(765611/765612).
[:CHANnel<n>]:SOURce[:CURRent]:PULSe:
BASE?
→ <current> The present current pulse base
value.
Example
:SOUR:CURR:PULS:BASE 0.75
:CHAN2:SOUR:CURR:PULS:BASE 5mA
:CHAN1:SOUR:CURR:PULS:BASE?
Description If CURRent is omitted, the GS820 assumes that
the present source function was specified.
[:CHANnel<n>]:SOURce[:CURRent]:ZERO:
IMPedance/?
Function Sets the impedance (high or low) for current zero
source or queries the current setting.
Syntax
[:CHANnel<n>]:SOURce[:CURRent]:ZERO:
IMPedance HIGH|LOW
HIGH Sets to high impedance.
LOW Sets to low impedance.
[:CHANnel<n>]:SOURce[:CURRent]:ZERO:
IMPedance?
→ HIGH
The current zero source is set to high
impedance.
→ LOW
The current zero source is set to low
impedance.
Example
:SOUR:CURR:ZERO:IMP LOW
:CHAN2:SOUR:CURR:ZERO:IMP HIGH
:CHAN1:SOOR:CURR:ZERO:IMP?
Description If CURRent is omitted, the GS820 assumes that
the present source function was specified.
16.2 Commands