4 Maintenance Procedures
90
March 14, 2014, 715004387 Rev. B
Removing and refitting the source enclosure
The optional combined APPI/APCI, NanoFlow, and TRIZAIC sources are
supplied as a complete source enclosure. To fit them, you must first remove
the standard source enclosure.
Removing the source enclosure from the instrument
Required material
Chemical-resistant, powder-free gloves
To remove the source enclosure:
1. Prepare the instrument according to the procedure on
,
“Preparing the instrument for operations on or inside its source”.
2. Remove the probe from the source.
•
If you are removing an ESI probe, see
•
If you are removing an IonSABRE II probe, see
3. Disconnect the probe adjuster and options cables from the instrument’s
connectors.
4. Pull the source enclosure release (located at the bottom, right-hand side)
outward, and swing open the enclosure.
Warning:
To avoid personal contamination with biohazardous
and/or toxic materials, always wear chemical-resistant,
powder-free gloves while performing this procedure. The source
components can be contaminated.
Warning:
To avoid burn injuries, take great care while working
with the probe and source; they can be hot.
Warning:
To avoid puncture wounds, take great care while
working with the source enclosure open if a corona pin is fitted.
The corona pin tip is sharp.
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...