Replacing the source assembly seals
March 14, 2014, 715004387 Rev. B
175
To remove the probe adjuster assembly probe and source enclosure seals:
1. Remove the source enclosure from the instrument (see
).
2. Use the O-ring removal kit to carefully remove the following seals from
the probe adjuster assembly (see
):
•
Probe seal
•
Nebulizer gas seal
3. Use the O-ring removal kit to carefully remove the following seals from
the source enclosure:
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Probe adjuster nebulizer gas seal
Probe adjuster assembly probe seal