Replacing the source assembly seals
March 14, 2014, 715004387 Rev. B
177
To fit the new source enclosure and probe adjuster assembly probe seals:
1. Ensure that all the grooves for seals are free from dirt and debris.
Tip:
If contamination is present, use 1:1 methanol/water, applied to a
lint-free cloth, to carefully clean the grooves.
Requirement:
Ensure that the tails of the source enclosure seals are
correctly located in the groove when fitting them to the source enclosure.
2. Fit the new source enclosure seal to the source enclosure.
Tip:
Start by feeding the seal into the groove at the bottom right-hand
corner, and then working around the seal in a counterclockwise
direction.
3. Fit the following new seals to the source enclosure:
•
Nebulizer gas seal
•
Desolvation gas seal
Requirement:
These seals have a special cross-section; fit them in the
groove as shown.
4. Fit the following new seals to the probe adjuster assembly:
•
Probe seal
•
Nebulizer gas seal
5. Refit the source enclosure to the instrument (see
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Groove
Seal
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...