Cleaning the ion block assembly
March 14, 2014, 715004387 Rev. B
127
2. Ensure that the isolation valve is closed.
3. Grasp the cone gas nozzle handle and use it to rotate the sampling cone
assembly through 90 degrees.
4. Slide the sampling cone assembly out of the ion block assembly.
Notice:
To ensure correct operation of the ion block assembly
after reassembly, follow these guidelines:
• Do not remove the sampling cone assembly retaining blocks.
• Do not adjust the screws securing the sampling cone assembly
retaining blocks.
Isolation valve handle
in closed position
Sampling cone
assembly retaining
blocks
Cone gas nozzle
handle