C Materials of Construction and Compliant Solvents
244
March 14, 2014, 715004387 Rev. B
Preventing contamination
For information on preventing contamination, refer to
Controlling
Contamination in LC/MS Systems
(part number 715001307). Visit
http://www.waters.com.
Items exposed to solvent
The items that appear in the following table can be exposed to solvent. You
must evaluate the safety issues if the solvents used in your application differ
from the solvents normally used with these items. See
for details
about the most common ingredients used to prepare mobile phases.
Items exposed to solvent:
Item
Material
Autotune reservoirs
High-density polyethylene
Corona discharge pin mounting
contact
PEEK™
Gas exhaust port
Aluminum
Gas tubes
Fluorinated ethylene propylene
Ion block
Stainless steel
Ion block support
PEEK
Isolation valve
Gold-plated aluminum/bronze
O-rings
Viton
®
or PTFE-encapsulated Viton
Probe adjuster bellows
PTFE/Viton
Probe adjuster assembly
Anodized aluminum, glass-filled
acetal, and stainless steel
Probe shaft
PEEK
Push-in gas fittings
Nickel/brass
Solvent waste/leak management
Tygon tubing
Source enclosure
Alochromed aluminum
Source enclosure view port
Toughened plate glass
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...