3 Changing the Mode of Operation
62
March 14, 2014, 715004387 Rev. B
ESCi mode
To run ESCi applications, you must fit an ESI probe and corona pin to the
ESI/APCI/ESCi source enclosure.
See “Installing the ESI probe” on
, “Installing the corona pin in the
source” on
, and “IntelliStart Fluidics System” on
Optimizing the ESI probe for ESCi operation
See the mass spectrometer’s online Help for details on how to optimize the ESI
probe for ESCi operation.
APCI mode
APCI mode, an option for the mass spectrometer, produces singly charged
protonated or deprotonated molecules for a broad range of nonvolatile
analytes.
The APCI interface consists of the ESI/APCI/ESCi enclosure fitted with a
corona pin and an IonSABRE II probe. Mobile phase from the LC column
enters the probe, where it is pneumatically converted to an aerosol, rapidly
heated, and vaporized or gasified at the probe tip.
APCI mode:
Hot gas from the IonSABRE II probe passes between the sample cone and the
corona pin, which is typically operated with a discharge current of 5 µA.
Mobile phase molecules rapidly react with ions generated by the corona
IonSABRE II probe
Sample cone
Corona pin
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...