Preparing the IntelliStart Fluidics system
March 14, 2014, 715004387 Rev. B
53
To install the reservoir bottles:
1. Remove the reservoir bottle caps.
2. Screw the reservoir bottles onto the instrument, as shown below.
3. For each reservoir bottle, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the bottle.
To install low-volume vials:
1. If a standard reservoir bottle is fitted, remove it.
2. Screw the low-volume adaptor into the manifold and tighten it.
3. Screw the low-volume vial into the adaptor.
4. For each low-volume vial, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the vial.
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while installing reservoir bottles.
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves while installing low-volume vials.
TP03410
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...