8. Route the open end of the exhaust tubing to a suitable exhaust vent.
9. Check the oil level in the pump (see “Checking the roughing pump oil
level” on
, and, if needed, “Adding oil to the roughing pump” on
Requirement:
To ensure correct operation of the roughing pump, do not
operate it when its oil level is less than 30% of the MAX level.
10. Connect electric cables to the roughing pump (see
).