![Waters Xevo TQD Overview And Maintenance Manual Download Page 173](http://html2.mh-extra.com/html/waters/xevo-tqd/xevo-tqd_overview-and-maintenance-manual_3711421173.webp)
Replacing the ion block source heater
March 14, 2014, 715004387 Rev. B
173
7. Use the needle-nose pliers to gently grasp the heat-shrink tubing on the
heater cartridge assembly and slide the assembly out of the ion block.
8. Dispose of the heater cartridge assembly.
9. Use the needle-nose pliers to gently grasp the heat-shrink tubing on the
new heater cartridge assembly and slide the assembly into the ion block.
10. Use the needle-nose pliers to position the 2 heater wire ring tags fully
down on the PEEK block terminals.
11. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 2 screws securing the heater wires to the PEEK
terminal block.
12. Fit the ion block cover plate onto the ion block assembly, and then use
the combined 2.5-mm Allen wrench and cone extraction tool to tighten
the 2 captive screws securing ion block cover plate.
13. Fit the ion block assembly onto the instrument (see
Notice:
To avoid damaging the heater cartridge assembly wires,
do not bend or twist them when fitting the assembly to the ion
block.
Notice:
To avoid a short circuit to the ion block cover, ensure that
the 2 heater cartridge ring tags are pushed fully down on the
PEEK block terminals.
Heat-shrink tubing
Heater cartridge assembly
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...