4 Maintenance Procedures
92
March 14, 2014, 715004387 Rev. B
Fitting the source enclosure to the instrument
Required material
Chemical-resistant, powder-free gloves
To fit the source enclosure:
1. Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
2. Close the source enclosure.
3. Connect the probe adjuster and options cables to the instrument’s
connectors.
Warning:
To avoid personal contamination with biologically
hazardous, or toxic materials, and to avoid spreading
contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when working with the
source components.
Warning:
To avoid puncture wounds, take great care while fitting
the source enclosure to the source if a corona pin is fitted (the pin
tip is sharp).
Notice:
To prevent the sprayer from colliding with the cone and,
consequently, breaking when you use a NanoFlow source, always
retract the stage before closing the source enclosure door.