4 Maintenance Procedures
174
March 14, 2014, 715004387 Rev. B
Replacing the source assembly seals
To avoid solvent vapors of biohazards and toxic materials leaking into the
laboratory atmosphere, replace the following seals at intervals not exceeding
one year:
•
Probe adjuster assembly probe seal
•
Probe adjuster assembly nebulization gas seal
•
Source enclosure seal
•
Nebulizer gas seal
•
Desolvation gas seal
Removing the probe adjuster assembly probe and source
enclosure seals
Required materials
•
Chemical-resistant, powder-free gloves
•
O-ring removal kit
Warning:
To avoid solvent vapors of biohazards and toxic
materials leaking into the laboratory atmosphere, replace the
seals listed below exactly as described in this section.
Warning:
To avoid solvent vapors of biohazards and toxic
materials leaking into the laboratory atmosphere, replace the
seals listed below at intervals not exceeding one year.