Replacing the IonSABRE II probe heater
March 14, 2014, 715004387 Rev. B
169
Fitting the new IonSABRE II probe heater
Required materials
•
Chemical-resistant, powder-free gloves
•
IonSABRE II probe heater
To fit the new IonSABRE II probe heater:
1. Use the probe adjuster knob to adjust the capillary so that it protrudes
approximately 0.5 mm from the end of the probe.
2. Carefully slide the probe heater over the capillary sleeve on the probe
assembly.
3. Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
4. Fit the probe to the instrument (see
5. In the Instrument Console, click API
to start the desolvation gas.
Notice:
To avoid damaging the probe heater’s electrical
connections, capillary sleeve, or capillary, take great care when
fitting the heater over the capillary sleeve.
Notice:
To avoid damaging the probe heater’s electrical
connections, do not twist the heater when fitting it to the probe
assembly.
Probe heater connections
Capillary sleeve
Capillary
0.5 mm
Summary of Contents for Xevo TQD
Page 2: ...ii March 14 2014 715004387 Rev B...
Page 22: ...xxii March 14 2014 715004387 Rev B...
Page 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Page 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Page 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Page 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Page 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...